HiScreen MabSelect PrismA
Home / Products / Chromatography Reagents / Antibody affinity chromatography /

HiScreen MabSelect PrismA

Product Information
Cat#No# Hi-008P
Product Overview This protein A column is prepacked with MabSelect PrismA chromatography resin. This affinity resin has been improved with an optimized high-flow agarose base matrix and a genetically engineered protein ligand, allowing efficient cleaning between runs. This allows future demands in monoclonal antibody purification to be met, including processing of many bispecific antibodies.
Description MabSelect PrismA is a next-generation Protein A chromatography resin that offers significantly enhanced alkaline stability and binding capacity for improved process economy in monoclonal antibody (mAb) processing. The resin builds on the proven track record of MabSelect and MabSelect SuRe resins in commercial mAb production. In comparison with its predecessors, however, MabSelect PrismA has been improved with an optimized highflow agarose base matrix and a genetically engineered Protein A-derived ligand, allowing future demands in mAb processing to be met
Characteristic Enhanced dynamic binding capacity compared with other protein A chromatography resins.
Excellent alkaline stability enables efficient cleaning and sanitization using 0.5–1.0 M NaOH.
Convenient method optimization and parameter screening enabled by the 10 cm bed height for efficient chromatography process development.
Maximum operating pressure 3 bar (0.3 MPa, 44 psi)
Ligand Coupling Method Single point attachment
Matrix Highly cross-linked agarose, spherical
Average particle size ~ 60 µm
Ligand Alkaline stabilized Protein A-derived (E. coli)
Coupling chemistry Epoxy
Dynamic binding capacity ~ 80 mg human IgG/mL resin at 6 min residence time. ~ 40 mg human IgG/mL at 2 min residence time.
Recommended flow rate < 600 cm/h
Recommended column height 100 mm
Chemical stability Stable to commonly used aqueous buffers in Protein A chromatography.
pH working range 3–12
CIP stability 2–14
Temperature stability 2°C to 40°C
Storage 2°C to 8°C in 20% ethanol
Shipping 20% ethanol On request 2% benzyl alcohol (BnOH).
Cleaning-in-place 1. Wash the column with 3 column volumes (CV) of start buffer.
2. Wash with 3 CV 0.5 to 1.0 M NaOH with a contact time of 15 minutes.
3. Wash immediately with at least 5 CV sterile and filtered start buffer at pH 7 to 8.
Pack size 1 × 4.7 mL
Maximum flow velocity 300 cm/h
Dimensions 7.7 × 100 mm
Column volume 4.7 ml
Column i.d. 7.7 mm
Column hardware pressure limit 0.8 MPa (8 bar)
Download Datasheet
For Research Use Only. Not intended for any clinical use. No products from Creative BioMart may be resold, modified for resale or used to manufacture commercial products without prior written approval from Creative BioMart.
Online Inquiry
Copyright © Creative BioMart. All rights reserved.