HiScreen Phenyl FF (High Sub)
Home / Products / Chromatography Reagents / Hydrophobic Interaction Chromatography /

HiScreen Phenyl FF (High Sub)

Cat# Product Name Size Price Qty Inquiry
Hi-144P HiScreen Phenyl FF (High Sub) Inquiry
Product Information
Cat#No# Hi-144P
Product Overview HiScreen Phenyl FF (high sub) columns are prepacked with Phenyl Sepharose 6 Fast Flow hydrophobic interaction chromatography (HIC) resins. The columns are an excellent choice for method optimization and parameter screening.
Description The small column volume of 4.7 mL and bed height of 10 cm make HiScreen columns excellent tools for method optimization, parameter screening, robustness testing, and convenient scale-up. Process fluid velocities can be applied, since the 10 cm bed height gives enough residence time and the results can then serve as basis for linear process scale-up.
Characteristic Phenyl Sepharose 6 Fast Flow is a well established, standard aromatic HIC resin for capture and intermediate purification requiring low to resin hydrophobicity.
Two levels of substitution help find optimal selectivity and binding capacity for a given application.
Prepacked, ready-to-use columns for convenient process development optimization.
10 cm bed height of HiScreen columns is designed to allow method optimization and parameter screening.
Easily connected in series to achieve 20 cm bed height.
Small bed volume gives fast results and minimal sample/buffer consumption.
Reproducible results, scalable to BioProcess columns packed with the same resin using the same linear fluid velocity.
Maximum operating pressure 1.5 bar [0.15 MPa] (22 psi)
Matrix Cross-linked agarose, 6%, spherical
Average particle size ~ 90 µm
Ligand Phenyl
Ligand density ~ 45 µmol/mL resin (high sub)
Recommended flow rate < 450 cm/h
Recommended column height 100 mm
Chemical stability Stable to commonly used aqueous buffers, 1.0 M NaOH5 , 8 M Urea, 6 M guanidine hydrochloride, 30% isopropanol, and 70% ethanol.
pH working range 3 to 13
CIP stability 2 to 14
Storage 4 to 30°C, 20% Ethanolor 0.01M NaOH
Cleaning-in-place 1. Wash with 3 column volumes (CV) of water or elution buffer.
2. Wash with 4 CV 0.5 to 1.0 M NaOH at a low flow rate (≤ 0.78 mL/min (≤ 100 cm/h)).
3. Wash with at least 3 CV water.
Pack size 4.7 mL
Maximum flow velocity 3.5 mL/min
Dimensions 7.7 × 100 mm
Column volume 4.7 mL
Column i.d. 7.7 mm
Column hardware pressure limit 0.8 MPa (8 bar, 116 psi)
Download Datasheet