HiTrap Capto Q ImpRes column
Home / Products / Chromatography Reagents / Ion Exchange Chromatography /

HiTrap Capto Q ImpRes column

Cat# Product Name Size Price Qty Inquiry
Hi-229P HiTrap Capto Q ImpRes column Inquiry
Product Information
Cat#No# Hi-229P
Product Overview HiTrap Capto Q ImpRes chromatography column is packed with a strong anion exchange modern resin for high resolution, small-scale protein purification.
Characteristic Suitable for screening selectivity, binding and elution conditions, and for small-scale purifications.
Quaternary ammonium (Q) strong anion exchanger.
Optimized for intermediate and polishing protein purification.
Convenient HiTrap format for easy connection to a syringe, peristaltic pump, or chromatography system.
Maximum operating pressure 5 bar [0.5 MPa] (70 psi)
Sample preparation 1. Dilution of cell paste: Add 5–10 ml of start buffer for each gram of cell paste.
2. Enzymatic lysis: 0.2 mg/ml lysozyme, 20 μg/ml DNAse, 1 mM MgCl2, 1 mM Pefabloc SC or PMSF (final concentrations). Stir for 30 min at room temperature or 4°C depending on the sensitivity of the target protein.
3. Mechanical lysis: Sonication on ice for approx. 10 min, homogenization with a French press or other homogenizer or freeze/thaw, repeated at least five times.
4. Adjust the pH of the lysate. The pH should be at least 0.5 units below (cation exchangers) or 0.5 units above (anion exchangers) the pI of the target molecule. Do not use strong bases or acids for pH-adjustment (precipitation risk). Apply the unclarified lysate on the column directly after preparation.
Metal ion capacity 0.15 to 0.18 mmol (Cl- ) /ml medium
Matrix High flow agarose
Particle Size 36 to 44 μm
Dynamic binding capacity >55 mg BSA /ml medium; >48 mg -lactoglobulin /ml medium.
Recommended column height 25 mm
Chemical stability All commonly used aqueous buffers, 1 M sodium hydroxide, 8 M urea, 6 M guanidine hydrochloride, 30% isopropanol and 70% ethanol.
pH working range 2 to 14
CIP stability 2 to 12
Storage 4 to 30°C, in 20% Ethanol
Cleaning-in-place 1. Wash with at least 2 column volumes (CV) of 2 M NaCl.
2. Wash with at least 4 CV of 1 M NaOH.
3. Wash with at least 2 CV of 2 M NaCl.
4. Rinse with at least 2 CV of distilled water.
5. Wash with 5 CV of start buffer for Capto Q ImpRes and Capto SP ImpRes or until eluent pH and conductivity have reached the required values.
Pack size 5 × 1 mL
Dimensions 7 × 25 mm
Column volume 1 mL
Column i.d. 7 mm
Column hardware pressure limit 5 bar (0.5 MPa)
Functional group -CH2N+(CH3)3
Download Datasheet