Product Information
Cat#No# Hi-232P
Product Overview HiTrap Capto S chromatography column is prepacked with a strong cation exchange modern resin for small scale protein purification.
Characteristic Suitable for screening of selectivity, binding and elution conditions, and small-scale purifications.
Sulfoethyl (S) strong cation exchanger.
Optimized for capture and intermediate protein purification.
Convenient HiTrap format for easy connection to a syringe, peristaltic pump, or chromatography system.
Maximum operating pressure 5 bar [0.5 MPa] (70 psi)
Sample preparation 1. Dilution of cell paste: Add 5–10 ml of start buffer for each gram of cell paste.
2. Enzymatic lysis: 0.2 mg/ml lysozyme, 20 μg/ml DNAse, 1 mM MgCl2, 1 mM Pefabloc SC or PMSF (final concentrations). Stir for 30 min at room temperature or 4°C depending on the sensitivity of the target protein.
3. Mechanical lysis: Sonication on ice for approx. 10 min, homogenization with a French press or other homogenizer or freeze/thaw, repeated at least five times.
4. Adjust the pH of the lysate. The pH should be at least 0.5 units below (cation exchangers) or 0.5 units above (anion exchangers) the pI of the target molecule. Do not use strong bases or acids for pH-adjustment (precipitation risk). Apply the unclarified lysate on the column directly after preparation.
Metal ion capacity 0.11 to 0.14 mmol H+/mL resin
Matrix Highly cross-linked agarose, spherical
Ionic Exchanger Type Strong cation, S
Average particle size ~ 90 μm
Dynamic binding capacity > 120 mg lysozyme/mL resin;> 60 mg ß-Lactoglobulin/mL resin.
Recommended flow rate 1 mL/min
Recommended column height 25 mm
Chemical stability Stable to commonly used aqueous buffers, 1.0 M NaOH, 8 M Urea, 6 M guanidine hydrochloride, 30% isopropanol and 70% ethanol.
pH working range 4 to 12
CIP stability 3 to 14
Storage 4 to 30°C, 0.2 M Sodium Acetate in 20% Ethanol.
Cleaning-in-place 1. Wash with at least 2 column volumes (CV) of 2 M NaCl.
2. Wash with at least 4 CV of 1 M NaOH.
3. Wash with at least 2 CV of 2 M NaCl.
4. Rinse with at least 2 CV of distilled water.
5. Wash with 5 CV of start buffer for Capto Q ImpRes and Capto SP ImpRes or until eluent pH and conductivity have reached the required values.
Pack size 5 × 1 mL
Maximum flow velocity 700 cm/h
Dimensions 7 × 25 mm
Column volume 1 mL
Column i.d. 7 mm
Column hardware pressure limit 5 bar (0.5 MPa, 72 psi)
Functional group -SO3 -