Cat# | Product Name | Size | Price | Qty | Inquiry |
---|---|---|---|---|---|
Hi-234P | HiTrap DEAE Sepharose FF | Inquiry |
Product Information | |
Cat#No# | Hi-234P |
Product Overview | HiTrap DEAE FF is prepacked with DEAE Sepharose Fast Flow, and is a weak anion exchanger for small-scale protein purifications as well as screening of binding and elutions conditions. |
Characteristic | Convenient and affordable for fast, easy ion exchange separations either alone or connected in series. The industry standard for ion exchange chromatography. High flow rates and good scale-up potential. Use a weak ion exchanger if the selectivity of a strong ion exchanger is insufficient. Predictable scale-up. |
Applications | For method scouting, group separations, sample concentration and sample clean-up of charged biomolecules. |
Maximum operating pressure | 5 bar [0.5 MPa] (70 psi) |
Sample preparation | The sample must be adjusted to the composition of the start buffer by buffer exchange using HiTrap Desalting, HiPrep 26/10 Desalting or PD-10 columns. The sample must be filtered through a 0.45 μm filter or centrifuged immediately before it is applied to the column. |
Metal ion capacity | 0.11–0.16 mmol Cl- /mL resin. |
Matrix | Cross-linked agarose, 6%, spherical. |
Ionic Exchanger Type | Weak anion |
Average particle size | ~ 90 μm |
Dynamic binding capacity | ~ 110 mg HSA/ mL resin. |
Recommended flow rate | < 4 ml/min |
Recommended column height | 25 mm |
Chemical stability | Stable to commonly used aqueous buffers, 1.0 M NaOH, 8 M urea, 6 M guanidine hydrochloride, 70% ethanol. |
pH working range | 2 to 12 |
CIP stability | 2 to 14 |
Storage | 4 to 30°C, 20% Ethanol |
Pack size | 5 × 1 mL |
Dimensions | 7 × 25 mm |
Column volume | 1 mL |
Column i.d. | 7 mm |
Column hardware pressure limit | 5 bar (0.5 MPa) |
Functional group | -N+(C2H5)2H |
Download Datasheet |
For Research or Industrial Raw Materials, Not For Personal Medical Use!
Online Inquiry